Blood sample handling best practices

Collection tubes Collection tubes Collection tubes

Fill blood tubes in the correct order to avoid contamination. EDTA contamination of chemistry samples may affect electrolyte results, and cause a falsely low Ca and falsely high K+.*1

Chemistry2,3

Whole blood samples must be inverted 8-10 times after collection and re-inverted just prior to use.

  • Never shake blood sample tubes.
  • 1.3 mL and smaller tubes may need additional inversions for proper mixing. It is recommended 10-15 inversions for smaller tubes.
  • Do not rely on a rocker to mix blood samples properly; rockers do not take the place of proper tube inversion.

Normal

Normal plasma and serum samples are straw colored, and do not have a yellow, red, or pink tinge.

Icteric

Icteric plasma and serum samples have a yellow color due to a disease or condition that causes excess bilirubin in the blood.

Unfortunately, nothing can be done to correct an icteric sample.

Hemolyzed

Hemolyzed plasma and serum samples have a pink/red tint due to ruptured red blood cells. Avoid hemolysis by using proper sample collection and handling techniques.1

Clotted

Clotted samples may have visible red clots that stick to wooden applicator sticks when swirled in a sample. Traumatic or delayed blood collection can lead to micro and/or macro clots.1

Avoid clotted samples by inverting blood tube appropriately immediately after filling. Re-draw a new sample if clots are present.

Lipemic

Lipemic plasma and serum samples have a milky appearance due to a high concentration of fat in the blood.

Avoid lipemia by using a fasted patient sample whenever possible.1 Remind clients to refrain from feeding their pets prior to their appointment.

Chemistry2

Lithium Heparin whole blood samples at room temperature must be run within 1 hour,6 or separated to serum* or plasma and run as soon as possible.7 Serum and plasma samples may be stored refrigerated§ for up to 48 hours.6

Discover the difference Zoetis Diagnostics can make in your clinic

Contact us today to learn more about how our diagnostics portfolio can help elevate your patient care.

Customer Service, Support & Ordering

Mon – Fri: 8:30am – 6:30pm ET

Vetscan and i-STAT Technical Support

24 hours a day, 7 days a week, 365 days a year

1-888-963-8471

Zoetis Reference Laboratories Customer Service, Support & Ordering

Mon – Fri: 8:00am – 9:30pm ET

Saturday: 9:00am – 8:30pm ET

Sunday: Closed

1-888-965-9652

Need Technical Support?

Our support team is available to answer product questions and provide guidance.||

* Always fill to manufacturer’s fill line.

Room Temperature (68-77 ̊F).

Stored plasma and serum samples must be separated and kept in a stoppered test tube containing no additive.

§ Refrigerated Temperature (36-46°F).

|| If you are a pet owner looking for treatment recommendations, please contact your veterinarian.

References: 1. Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - overfilling the tube - lead to small clots (p.2)
Conversely, insufficient EDTA in relation to blood will lead to clot formation. Small clots in the sample, which might be missed when visual; Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - Blood tube fill order (p.2) Under- or over-filling of blood tubes: tubes should be filled to the correct volume and gently inverted to mix the blood with the pre- measure; Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - Under-filled EDTA tube (p.2) If the concentration of EDTA is excessive in relation to blood volume (tube under-filling ), cells will shrink and falsely lower the PCV. E 2. Legacy Abaxis VetScan VS2 Operator's Manual (v2.0) 3. Weiser, G., Veterinary Hematology and Clinical Chemistry, 2012, 3-33 (v2.0) 4. Wu, D, et al., EBioMedicine, 2017 (v1.0) 5. Kitchens, JL., Maryville College, 2006. (v1.0) 6. Weiser, G. Sample Collection, Processing, and Analysis of Laboratory Service Options [Chapter 2]. Veterinary Hematology and Clinical Chemistry, 2012: p.3. 7. Monti P, Archer J. Quality Assurance and Interpretation of Laboratory Data [Chapter 2]. BSAVA Manual of Canine and Feline Clinical Pathology. 3rd ed.; 2016: p.13.

Request support

Request Support

Use this form to make product inquiries, report a product related event, or raise a product quality issue, and a Zoetis Representative will be in touch.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close

Request a demonstration

Request More Information

Ready to see our diagnostic portfolio in action? Simply complete the form below and a Zoetis representative will be in touch to answer your questions, provide more information or schedule your complimentary demonstration.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close

Request more information

Request More Information

Interested in hearing more about the new technology partnership with Covetrus Pulse™? Simply complete the form below to learn more or schedule an installation with a Zoetis Representative.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close

Request an Integration

Request an Integration

Use this form to request a bidirectional integration and a Zoetis representative will be in touch.

All fields are required.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close