Blood sample handling best practices

Collection tubes Collection tubes Collection tubes

Fill blood tubes in the correct order to avoid contamination. EDTA contamination of chemistry samples may affect electrolyte results, and cause a falsely low Ca and falsely high K+.*1

Tube-Handling-min Tube-Handling-min Tube-Handling-min

Chemistry2,3

Whole blood samples must be inverted 8-10 times after collection and re-inverted just prior to use.

  • Never shake blood sample tubes.
  • 1.3 mL and smaller tubes may need additional inversions for proper mixing. It is recommended 10-15 inversions for smaller tubes.
  • Do not rely on a rocker to mix blood samples properly; rockers do not take the place of proper tube inversion.
Normal-Sample-Quality-min

Normal

Normal plasma and serum samples are straw coloured, and do not have a yellow, red or pink tinge.

Normal-Sample-Quality-Icetric

Icteric

Icteric plasma and serum samples have a yellow colour due to a disease or condition that causes excess bilirubin in the blood.

Unfortunately, nothing can be done to correct an icteric sample.

hemolysis-min

Haemolysed

Haemolysed plasma and serum samples have a pink/red tint due to ruptured red blood cells. Avoid haemolysis by using proper sample collection and handling techniques.1

Clotted-Sample-Quality-min

Clotted

Clotted samples may have visible red clots that stick to wooden applicator sticks when swirled in a sample. Traumatic or delayed blood collection can lead to micro and/or macro clots.1

Avoid clotted samples by inverting blood tube appropriately immediately after filling. Re-draw a new sample if clots are present.

lipemia-min

Lipaemic

Lipaemic plasma and serum samples have a milky appearance due to a high concentration of fat in the blood.

Avoid lipaemia by using a fasted patient sample whenever possible.1 Remind clients to refrain from feeding their pets prior to their appointment.

VS2-circle VS2-circle VS2-circle

Chemistry2

Lithium Heparin whole blood samples at room temperature must be run within 1 hour,6 or separated to serum* or plasma and run as soon as possible.7 Serum and plasma samples may be stored refrigerated§ for up to 48 hours.6

Discover the difference Zoetis Diagnostics can make in your clinic

Contact us today to learn more about how our diagnostics portfolio can help you provide elevated patient care.

Customer Service and Support

Monday - Friday

8:30 am - 5:00 pm

nzcontactus@zoetis.com

0800 ZOETIS (963 847)

Zoetis Diagnostic Technical Support

0800 100 109 (option 3)

Zoetis Reference Laboratories Customer Service, Support & Ordering

Mon – Fri: 8:00am – 9:30pm ET

Saturday: 9:00am – 8:30pm ET

Sunday: Closed

1-888-965-9652

Need Technical Support?

Our support team is available to answer product questions and provide guidance.||

* Always fill to manufacturer’s fill line.

Room Temperature (20-25 ̊C).

Stored plasma and serum samples must be separated and kept in a stoppered test tube containing no additive.

§ Refrigerated Temperature (2-7°F).

|| If you are a pet owner looking for treatment recommendations, please contact your veterinary surgeon.

References: 1. Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - overfilling the tube - lead to small clots (p.2)
Conversely, insufficient EDTA in relation to blood will lead to clot formation. Small clots in the sample, which might be missed when visual; Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - Blood tube fill order (p.2) Under- or over-filling of blood tubes: tubes should be filled to the correct volume and gently inverted to mix the blood with the pre- measure; Monti, P, Archer, J, BSAVA Manual of Canine and Feline Clinical Pathology, 2016, 11-26 (v2.0) - Under-filled EDTA tube (p.2) If the concentration of EDTA is excessive in relation to blood volume (tube under-filling ), cells will shrink and falsely lower the PCV. E 2. Legacy Abaxis VetScan VS2 Operator's Manual (v2.0) 3. Weiser, G., Veterinary Haematology and Clinical Chemistry, 2012, 3-33 (v2.0) 4. Wu, D, et al., EBioMedicine, 2017 (v1.0) 5. Kitchens, JL., Maryville College, 2006. (v1.0) 6. Weiser, G. Sample Collection, Processing, and Analysis of Laboratory Service Options [Chapter 2]. Veterinary Haematology and Clinical Chemistry, 2012: p.3. 7. Monti P, Archer J. Quality Assurance and Interpretation of Laboratory Data [Chapter 2]. BSAVA Manual of Canine and Feline Clinical Pathology. 3rd ed.; 2016: p.13.

Request support

Request Support

Use this form to make product inquiries, report a product related event, or raise a product quality issue, and a Zoetis Representative will be in touch.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close

All fields are required.

This field is required

This field is required

This field is required

This field is required Please enter a valid e-mail address

This field is required Please enter a valid phone number

Please select at least one testing application

This field is required

Please accept the terms of use and privacy policy to continue

Request a demonstration

Request more information

Ready to see our diagnostic portfolio in action? Simply complete the form below and a Zoetis representative will be in touch to answer your questions and provide more information.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

CLOSE

Sorry, form submission has failed. Please try again.

CLOSE

Request more information

Request More Information

Interested in hearing more about the new technology partnership with Covetrus Pulse™? Simply complete the form below to learn more or schedule an installation with a Zoetis Representative.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close

Request an Integration

Request an Integration

Use this form to request a bidirectional integration and a Zoetis representative will be in touch.

All fields are required.

Thank you for your submission!

Your Zoetis representative will reach out to you soon.

close

Sorry, form submission has failed. Please try again.

close